RALCO - manual collimator models
R72
Single-layer square field model for installation on portable and mobile units with fixed anode x-ray tubes. Plastic ABS cover. Operation with rotating anode x-ray tube is subject to technical assessment!
Technical data:
- Mounting plane at 70 mm from the x-ray focus
- Maximum radiation protection 125 kVp – 4 mA
- Two knobs on either side of the unit provide for the manual control of shutters
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min. 0 x 0 cm to max. 43 x 43 cm at 100 cm FFD (SID)
R72S
Single-layer square field manual collimator, square-field, single-layer, compact and lightweight. Plastic cover. Ideal for installation on portable x-ray equipment.
Technical data:
- Mounting plane at 52 mm from the x-ray focus
- Maximum radiation protection: 125 kVp - 4 mA
- Two knobs on either side of the unit provide for the manual control of shutters
- High luminosity provided by a white power LED (power supply 1 A)
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min. 0 x 0 cm to max. 43 x 43 cm at 100 cm FFD (SID)
R104
Single-layer square field model for installation on mobile units with fixed anode x-ray tubes. Steel covers. Operation with rotating anode x-ray tube is subject to technical assessment!
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 125 kVp – 4 mA
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min. 0 x 0 cm to max. 43 x 43 cm at 100 cm FFD (SID)
R104/A
Single-layer square field model for installation on mobile units with fixed anode x-ray tubes. Plastic covers. Operation with rotating anode x-ray tubes requires the adoption an optional item RO 278.
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 125 kVp – 4 mA
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- - Field size variable from min. 0 x 0 cm to max. 43 x 43 cm at 100 cm FFD (SID)
R104 F
Single-layer square field model for installation on mobile units with fixed anode x-ray tubes. Plastic covers. R104F is designed to accommodate additional filtration (optional item RO 319). Operation with rotating anode x-ray tubes requires the use of optional item RO 278.
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 125 kVp – 4 mA
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Optional manual positioning filter
- X-ray field simulated by light field provided by a white power LED
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min: 0 x 0 cm to max: 43 x 43 cm at 100 cm FFD (SID)
R105
Single-layer square field model for installation on mobile units with fixed anode x-ray tubes. Steel covers. Operation with rotating anode x-ray tube is subject to technical assessment!
Technical data:
- Mounting plane at 100 mm from the x-ray focus
- Maximum radiation protection 125 kVp – 4 mA
- Two pairs of moving shutters (near the exit window) and a fix cone (near the focus)
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min: 0 x 0 cm to max: 43 x 43 cm at 100 cm FFD (SID)
R108
Multi-layer square field model for installation on stationary and mobile units with fixed or rotaring anode x-ray tubes. Plastic covers.
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 125 kVp – 4 mA (150 kVp optional)
- Six pairs of shutters in 3 layers to accurately define the X-ray field edges
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- External adjustment of mirror angulation
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min: 0 x 0 cm to max: 43 x 43 cm at 100 cm FFD (SID)
R108+RO288
Multi-layer square field model for installation on mobile units with fixed or rotaring anode x-ray tubes. Power LED for the light-field.
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 125 kVp – 4 mA (150 kVp optional)
- Six pairs of shutters in 3 layers to accurately define the X-ray field edges
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- High luminosity provided by a solid state power LED
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- External adjustment of mirror angulation
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min: 0 x 0 cm to max: 43 x 43 cm at 100 cm FFD (SID).
R108 F
Multi-layer square field model for installation on stationary and mobile units with fixed or rotaring anode x-ray tubes. Power LED for the light-field. R108F is designed to accommodate additional filtration (optional item RO 319).
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 125 kVp – 4 mA (150 kVp optional)
- Six pairs of shutters in 3 layers to accurately define the X-ray field edges
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Optional manual positioning filter
- High luminosity provided by a solid state power LED
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- External adjustment of mirror angulation
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min: 0 x 0 cm to max: 43 x 43 cm at 100 cm FFD (SID)
R221/A
Multi-layer square field model for installation on stationary and mobile units with fixed or rotaring anode x-ray tubes. Plastic covers.
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 150 kVp – 4 mA
- Six pairs of shutters in 3 layers to accurately define the X-ray field edges
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- External adjustment of mirror angulation
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min: 0 x 0 cm to max: 43 x 43 cm at 100 cm FFD (SID)
R302/A
Multi-layer square field model for installation on stationary units with fixed or rotaring anode x-ray tubes. Plastic covers.
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 150 kVp – 4 mA
- Six pairs of shutters in 3 layers to accurately define the X-ray field edges
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- External adjustment of mirror angulation
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min: 0 x 0 cm to max: 43 x 43 cm at 90 cm FFD (SID)
R302L/A
Multi-layer square field model for installation on stationary units with fixed or rotaring anode x-ray tubes. Plastic covers. This collimator features a laser line for Potter Bucky/collimator alignment.
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 150 kVp – 4 mA
- Six pairs of shutters in 3 layers to accurately define the X-ray field edges
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- External adjustment of mirror angulation
- Minimum inherent filtration: 2 mm aluminium equivalent (1mm on request)
- Field size variable from min: 0 x 0 cm to max: 43 x 43 cm at 90 cm FFD (SID)
R302F/A
Multi-layer square field model for installation on stationary units with fixed or rotaring anode x-ray tubes. Plastic covers. This collimator features a laser line for Potter Bucky/collimator alignment.
Technical data:
- Mounting plane at 80 mm from the x-ray focus
- Maximum radiation protection 150 kVp – 4 mA
- Six pairs of shutters in 3 layers to accurately define the X-ray field edges
- Two knobs on the front panel of the collimator provide for the manual control of shutters
- Three filters on rotating support - manual positioning of filters
- Electronic 30 sec timer for the activation of the light-field simulating the x-ray field
- External adjustment of mirror angulation
- Minimum inherent filtration: 2 mm aluminium equivalent (0.3mm or 1mm on request)
- Field size variable from min: 0 x 0 cm to max: 43 x 43 cm at 90 cm FFD (SID)
News
ECR 2017
Pascal Team has exhibited it’s complete range of digital X-ray systems in ECR2017, the largest radiological meeting in Europe. More than 20.000 participants from over 100 countries visited the scientific and educational programme, and the related technical exhibition this year. Exhibitors from all over the world presented the results of their latest researches and developments.
The inspiration of our company name
Blaise Pascal
Blaise Pascal, the prominent French scientist was born in Clemont-Ferrand in 1623. He contributed significally in the fields of mathematics, physics, philosophy and theology as well.
In honor of his intellectual curiosity and diligence, assiduous and successful simultaneous work in different areas of science, we treat him as our model and have chosen his name for our company. ....